This document specifies the design and dimensions for two small-bore connectors intended to be used for connections in respiratory
applications of
medical devices and
accessories. One
connector (R1) is intended for use on
medical devices and
accessories subjected to pressures up to 15 kPa (e.g. a
breathing system). The other
connector (R2) is intended for use on
medical devices and
accessories subjected to higher pressures between 15 kPa and 600 kPa (e.g. oxygen therapy tubing).
NOTE 1: The pressure is related to pressure available at the source to which the
medical device is connected.
NOTE 2: The intended
application does not preclude the use of other
connectors on
medical devices or
accessories within this
application.
NOTE 3: Requirements for alternative connectors for this intended application are specified in ISO 80369-1.
This document does not specify requirements for the
medical devices or
accessories that use these
connectors. Such requirements are given in device-specific standards.
NOTE 4: If a device-specific standard does not exist, the performance and material requirements specified in ISO 80369-1 can be used as guidance.