Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use - Part 2-091: Particular Requirements for Cabinet X-Ray Systems
UL 61010-2-091 (IEC 61010-2-091:2012)
1 Scope and object
This clause of Part 1 is applicable, except as follows:
1.1.1 Equipment included in scope
Replacement:Replace the text with the following:A CABINET X-RAY SYSTEM is a system that contains an X-ray tube installed in a cabinet which, independently of existing architectural structures except the floor on which it may be placed, is intended to contain at least that portion of a material being irradiated, provide radiation attenuation, and exclude personnel from the interior during generation of X-radiation.These CABINET X-RAY SYSTEMS are used in industrial, commercial, and public environments, for example, to inspect materials, to analyze materials, and to screen baggage.

1.1.2 Equipment excluded from scope
Addition:Add the following new items to the list:
aa) equipment intended to apply X-radiation to humans or animals;
bb) equipment incorporating an X-ray tube but not incorporating complete shielding against X-radiation hazards, such as:
  • equipment intended to be used within a shielded room which excludes personnel during operation;
  • equipment intended to be used with separate portable or temporary shielding;
  • equipment intended to produce an emerging beam of X-radiation.
1.2.1 Aspects included in scope
Addition:Add the following text at the end of the first paragraph:This part of IEC 61010 specifies requirements for the design and methods of construction of CABINET X-RAY SYSTEMS to provide adequate protection for OPERATORS, bystanders, trained service personnel, and the surrounding area against unintentionally-emitted X-radiation and from mechanical HAZARDS related to their conveyors.
Edition : 1
Published : 08/08/2014

History

UL 61010-2-091
Published Date: 07/31/2019
Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-091: Particular requirements for cabinet X-Ray systems
$194.7
UL 61010-2-091
Published Date: 08/08/2014
Safety requirements for electrical equipment for measurement, control, and laboratory use - Part 2-091: Particular requirements for cabinet X-Ray systems

Related products


Best-Selling Products

SPI AQ-103
Published Date: 01/01/1994
Cosmetic Specifications of Injection Molded Parts
SPI AR-101
Published Date: 10/01/1996
Customs and Practices of the Moldmaking Industry
SPI AU 123
Published Date: 01/01/1992
Visual Evaluation of Reflected Color (Formerly PB-1)
SPI B151.1-1997
Published Date: 01/01/1997
Plastics Machinery - Horizontal Injection Molding Machines - Safety Requirements for Manufacture, Care, and Use
$11.1
SPI B151.15-2003
Published Date: 01/01/2003
Extrusion Blowmolding Machines - Safety Requirements for the Manufacture Care and Use
$13.8
SPI B151.2-1999
Published Date: 01/01/1999
Film Casting Machines - Construction, Care, and Use (revision and redesignation of ANSI B151.2-1982 (R1988))
$9.9