UL 61010A-2-061 PDF

UL 61010A-2-061 PDF

Name:
UL 61010A-2-061 PDF

Published Date:
04/10/2002

Status:
[ Withdrawn ]

Description:

Electrical Equipment for Laboratory Use: Part 2: Particular Requirements for Laboratory Atomic Spectrometers with Thermal Atomization and Ionization

Publisher:
Underwriters Laboratories

Document status:
Active

Format:
Electronic (PDF)

Delivery time:
10 minutes

Delivery time (for Russian version):
200 business days

SKU:

Choose Document Language:
Need Help?
Electrical Equipment for Laboratory Use; Part 2: Particular Requirements for Laboratory Atomic Spectrometers with Thermal Atomization and Ionization

UL 61010A-2-061

1 Scope and object

This clause of part 1 is applicable except as follows:

1.1 Scope

Replacement:Replace the complete text by the following:This International Standard applies to electrically powered laboratory atomic spectrometers with thermal atomization.NOTE - Examples include atomic absorption spectrometers, emission flame photometers, atomic fluorescence spectrophotometers, inductively coupled plasma spectrometers, microwave coupled plasma spectrometers, and mass spectrometers, all with thermal atomization and ionization (including tubing and connectors which are provided by the manufacturer for connection to external supplies).

1.1.2 Equipment excluded from scope

Addition:Add the following new dash:

- thermal atomization detectors (flame ionization detectors) used in gas chromatography.
Edition : 1st
Number of Pages : 26
Published : 04/10/2002

History


Related products


Best-Selling Products