Description:

Standard Practice for Evaluating the Performance of Respirable Aerosol Samplers

Publisher:
ASTM International

Document status:
Active

Format:
Electronic (PDF)

Delivery time:
10 minutes

Delivery time (for Russian version):
200 business days

SKU:

Choose Document Language:
$18
Need Help?

1.1 This practice covers the evaluation of the performance of personal samplers of non-fibrous respirable aerosol. The samplers are assessed relative to a specific respirable sampling convention. The convention is one of several that identify specific particle size fractions for assessing health effects of airborne particles. When a health effects assessment has been based on a specific convention it is appropriate to use that same convention for setting permissible exposure limits in the workplace and ambient environment and for monitoring compliance. The conventions, which define inhalable, thoracic, and respirable aerosol sampler ideals, have now been adopted by the International Standards Organization (Technical Report ISO TR 7708), the Comit Europen de Normalisation (CEN Standard EN 481), and the American Conference of Governmental Industrial Hygienists (ACGIH, Ref (1)), developed (2) in part from health-effects studies reviewed in Ref (3) and in part as a compromise between definitions proposed in Refs (3,4).

1.2 This practice is complimentary to Test Method D 4532, which specifies a particular instrument, the 10-mm cyclone. The sampler evaluation procedures presented in this practice have been applied in the testing of the 10-mm cyclone as well as the Higgins-Dewell cyclone., Details on the evaluation have been recently published (5-7) and can be incorporated into revisions of Test Method D 4532.

1.3 A central aim of this practice is to provide information required for characterizing the uncertainty of concentration estimates from samples taken by candidate samplers. For this purpose, sampling accuracy data from the performance tests given here can be combined with information as to analytical and sampling pump uncertainty obtained externally. The practice applies principles of ISO GUM, expanded to cover situations common in occupational hygiene measurement, where the measurand varies markedly in both time and space. A general approach (8) for dealing with this situation relates to the theory of tolerance intervals and may be summarized as follows: Sampling/analytical methods undergo extensive evaluations and are subsequently applied without re-evaluation at each measurement, while taking precautions (for example, through a quality assurance program) that the method remains stable. Measurement uncertainty is then characterized by specifying the evaluation confidence (for example, 95 %) that confidence intervals determined by measurements bracket measurand values at better than a given rate (for example, 95 %). Moreover, the systematic difference between candidate and idealized aerosol samplers can be expressed as a relative bias, which has proven to be a useful concept and is included in the specification of accuracy (3.2.3-3.2.10).

1.4 Units of the International System of Units (SI) are used throughout this practice and should be regarded as standard.

This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.


File Size : 1 file , 150 KB
Note : This product is unavailable in Russia, Ukraine, Belarus
Number of Pages : 9
Published : 04/01/2007
Redline File Size : 2 files , 310 KB

History

ASTM D6061-01(2018)e1
Published Date: 12/01/2018
Standard Practice for Evaluating the Performance of Respirable Aerosol Samplers
$14.4
ASTM D6061-01(2012)e1
Published Date: 04/01/2012
Standard Practice for Evaluating the Performance of Respirable Aerosol Samplers
$18
ASTM D6061-01(2007)e1
Published Date: 04/01/2007
Standard Practice for Evaluating the Performance of Respirable Aerosol Samplers
$18
ASTM D6061-01
Published Date: 10/10/2001
Standard Practice for Evaluating the Performance of Respirable Aerosol Samplers
$18
ASTM D6061-96
Published Date: 01/01/1996
Standard Practice for Evaluating the Performance of Respirable Aerosol Samplers
$17.4

Related products

ASTM D4298-22
Published Date: 09/01/2022
Standard Guide for Intercomparing Permeation Tubes to Establish Traceability
$14.4
ASTM D3686-20
Published Date: 03/01/2020
Standard Practice for Sampling Atmospheres to Collect Organic Compound Vapors (Activated Charcoal Tube Adsorption Method)
$15
ASTM D4856-23
Published Date: 09/01/2023
Standard Test Method for Determination of Sulfuric Acid Mist in Workplace Atmospheres Collected on Mixed Cellulose Ester Filters (Ion Chromatographic Analysis)
$16.5
ASTM D5466-21
Published Date: 08/14/2021
Standard Test Method for Determination of Volatile Organic Compounds in Atmospheres (Canister Sampling, Mass Spectrometry Analysis Methodology)
$22.5

Best-Selling Products

CTI 105-ATC (00)
Published Date: 02/01/2000
CTI Code Tower Standard Specifications for Acceptance Test Code for Water Cooling Towers
$31.2
CTI 105 S (11)
Published Date: 07/01/2011
Supplement to CTI Code Tower Standard Specifications for Acceptance Test Code for Closed Circuit Cooling Towers
$10.5
CTI 105 S (96)
Published Date: 07/01/1996
CTI Code Tower Standard Specifications for Acceptance Test Code for Closed Circuit Cooling Towers
$10.5
CTI 106-ATC (04)
Published Date: 07/01/2004
Acceptance Test Code for Mechanical Draft Evaporative Vapor Condensers
$10.5
CTI ATC-105 (19)
Published Date: 02/01/2019
Acceptance Test Code For Cooling Towers
$37.5
CTI ATC-106 (11)
Published Date: 07/01/2011
Acceptance Test Code for Mechanical Draft Evaporative Vapor Condensers
$7.5